True RMS meter

 True RMS meter

There exists a fundamental difference between the readings on a normal ac meter and on a true rms meter. The first uses a D' Arsonval movement with a full or half wave rectifier, and averages the values of the instantaneous rectified current.
The rms meter, however, averages the squares of the instantaneous current values (proportional, for example, to the instantaneous heating effect). The scale of the true rms meter is calibrated in terms of the square roots of the indicated current values. The resulting reading is therefore the square root of the  average of the squared instantaneous input values, which is the rms value of the measured alternating current.
A true rms meter is always a combination of a normal mean value indicating meter and a squaring device whose output at any instant is proportional to the instantaneous squared input. 
It can be shown that the ac component of the voltage developed across the common collector resistors of two transistors that are connected in parallel, and between the bases of which a small ac voltage is applied, is proportional to the square of the applied input voltage.
Squaring device


The basic circuit of above Fig. employing two transistors is completed by a bridge arrangement in which the dc component is cancelled out. This bridge arrangement is given in Fig.
True RMS meter


One side of the bridge consists of two parallel connected transistors Q2 and Q3, and a common collector resistor R13. The side of the bridge, employing P1 for bias setting, is the basic squaring circuit. The other side of the bridge is made of transistor Q4 (whose base is biased by means of potentiometer P2 and collector resistance R16.)
Potentiometer P1,base bias balance of the squaring circuit, must be adjusted for the symmetrical operation of transistors Q2 and Q3. To do this, the polarity of a small dc input voltage applied to terminals A and B (bases of Q2 and Q3) has to be reversed, and the reading of the output meter must be the same for both input polarities.

Potentiometer P2 must be set so that for zero input signal (terminals A and B short-circuited), the bridge is balanced and the meter reads zero. The balance condition is reached if the voltage drop across the collector resistance R13 of Q2 – Q3, and collector resistance R16 of Q4, are equal.

Transistor Q1 is used to improve the temperature stability of the whole circuit, which is basically obtained by the emitter resistance R10. Optimum temperature compensation is obtained if the voltage drop across the emitter resistance for no signal is 0.7 V for the silicon transistor.

The low current through Q2, Q3, Q4 requires a large emitter resistance value to fulfi l the condition for compensation. Therefore, another transistor, Q1 has been added to compensate for the temperature changes of Q2 and Q3. 
The bias on this transistor has to be adjusted by selecting appropriate values of R 8 and R9 so that the voltage drop across R10 in the balanced condition is 0.7 V for silicon transistor. The input of the squaring devices (AB) is connected to a voltage divider that is calibrated in seven ranges, namely 0.3, 1, 3, 10, 30, 100, and 300 volts.





Comments

Popular posts from this blog

Concepts and types of Digital voltmeter

Digital Voltmeter (DVM) Definition : Digital Voltmeter is a  voltage-sensitive device . It  measures AC  or  DC voltage  and  displays  the value directly in  numeric form  instead of pointer deflection.  DVM  is an acronym for  Digital Voltmeter . DVM was first invented in  1954  by  Andrew Kay . There exist many factors that affect the measurement accuracy of a digital voltmeter(DVM). These are basically temperature, input impedance, variation in power supply voltage etc. As we know that an  analogue instrument  provides results by  pointer deflection  on a continuous scale. On contrary, a  digital instrument  provides results as  discrete numerals . Thus providing accuracy and versatility together. The input range of DVM may vary from  ±1 V  to  1000 V . Precision DVM offers  input resistance  of  1 GΩ  or high for a voltage range of less th...

DSO & MSO (Block Diagram)

DIGITAL STORAGE OSCILLOSCOPE (DSO) The digital storage oscilloscope eliminates the disadvantages of the analog storage oscilloscope. It replaces the unreliable storage method used in analog storage scopes with digital storage with the help of memory. The memory can store data as long as required without degradation. It also allows the complex processing of the signal by the high-speed digital signal processing circuits.  In this digital storage oscilloscope, the waveform to be stored is digitised and then stored in digital memory. The conventional cathode ray tube is used in this oscilloscope hence the cost is less. The power to be applied to memory is small and can be supplied by small battery. Due to this the stored image can be displayed indefinitely as long as power is supplied to memory. Once the waveform is digitised then it can be further loaded into the computer and can be analysed in detail. Block Diagram As done in all the oscilloscopes, the input signal is applied to the...

Maxwell’s inductance bridge

 Maxwell's Bridge Maxwell's bridge can be used to measure inductance by comparison either with variable standard self-inductance or with a standard variable capacitance. These measurements can be done by using Maxwell's Bridge in two different forms. Maxwell's Inductance Bridge Using this bridge, We can measure inductance by comparing it with a Standard variable self-inductance arranged in a bridge circuit as shown in Fig.  (a). Consider Maxwell's inductance bridge as shown in Fig  (a). Two branches consist of non-inductive resistances R1 and R2. One of the arms consists of Variable inductance with series resistance r. The remaining arm consists of unknown inductance Lx. Maxwell's Inductance Capacitance Bridge Using this bridge, we can measure inductance by comparing it with a variable standard capacitor. The bridge circuit diagram is shown in Fig. One of the ratio arms consists of resistance and capacitance in parallel. Hence it is simple to write the bridge eq...