Megger

 Resistances of the order of 0.1 M Ohm and upwards are classified as high resistances. These high resistances are measured by a portable, instrument known as a megger. It is also used for testing the insulation resistance of cables.

Principle of Operation

It is based on the principle of electromagnetic induction. The Figure shows the construction of a megger.


When a current-carrying conductor is placed in a uniform magnetic field it experiences a mechanical force whose magnitude depends upon the strength of the current and magnetic field. While its direction depends on the direction of the current and magnetic field.

Construction

It consists of a permanent magnet which provides the field for both Generator G and an ohmmeter. The moving element of the ohmmeter consists of three coils viz. current or deflection coil, pressure or control coil and compensating coil.  These coils are mounted on a central shaft which is free to rotate over a stationary C-shaped iron core. 
The coils are connected to the circuit through flexible leads called ligaments which do not produce a restoring torque on the moving element, consequently, the moving element takes up any position over the scale when the generator handle is stationary.

The current coil is connected in series with resistance R1 between one generator terminal and the test terminal T2. The series resistance R1 protects the current coil in the event of the test terminals getting short-circuited and also controls the range of the instrument. The pressure coil, in series with a compensating coil and protection resistance R2 is connected across the generator terminals. The compensating coil is included in the circuit to ensure better scale proportions. The scale is calibrated reversely means the normal position of the pointer indicates infinity while full-scale deflection indicates zero resistance.

Working
When the current flows from the generator, through the pressure coil, the coil tends to set itself at right angles to the field of the permanent magnet.
When the test terminals are open, corresponding to infinite resistance, no current flows through deflection coil. Thus the pressure coil governs the motion of the moving element making it move to its extreme anticlockwise position. The pointer comes to rest at the infinity end of the scale. 
When the test terminals are short-circuited i.e. corresponding to zero resistance, the current from the generator flowing through the current coil is large enough to produce sufficient torque to overcome the counter-clockwise torque of the pressure coil. Due to the pointer moving over a scale showing zero resistance. 
When the high resistance to be tested is connected between terminals Tl and T2 the opposing torques of the coils balance each other so that pointer attains a stationary Position at some intermediate point on scale. The scale is calibrated in megaohms so that resistance is directly indicated by the pointer.

The guard ring is provided to eliminate the error due to leakage current. The supply meter is usually given by a hand-driven permanent magnet d.c. generator sometimes motor-driven generator may also be used.

Applications

The megger can be used to determine whether there is sufficiently high resistance between the conducting part of a circuit and the ground. This resistance is insulation resistance.
The megger can also be used to test continuity between any two points. When connected to the two points, if the pointer shows full deflection then there is a continuity between them.

Comments

Popular posts from this blog

Analog to Digital Converters: Successive Approximation A/D Converter

  Analog to Digital Converter & Its Working Almost every environmental measurable parameter is in analog form like temperature, sound, pressure, light, etc. Consider a temperature  monitoring system  wherein acquiring, analyzing, and processing temperature data from sensors is not possible with digital computers and processors. Therefore, this system needs an intermediate device to convert the analog temperature data into digital data in order to communicate with digital processors like microcontrollers and microprocessors. Analog to Digital Converter (ADC) is an electronic integrated circuit used to convert the analog signals such as voltages to digital or binary form consisting of 1s and 0s. Most of the ADCs take a voltage input as 0 to 10V, -5V to +5V, etc., and correspondingly produces digital output as some sort of a binary number. What is Analog to Digital Converter? A converter that is used to change the analog signal to digital is known as an analog to digital...

Single Channel Vs Multi Channel DAS

  Single Channel Data Acquisition System: A Single Channel Data Acquisition System consists of a signal conditioner followed by an analog to digital (A/D) converter, performing repetitive conversions at a free running, internally determined rate. The outputs are in digital code words including over range indication,  polarity  information and a status output to indicate when the output digits are valid. A Single Channel Data Acquisition System is shown in Fig. 17.3. The digital outputs are further fed to a storage or printout device, or to a digital computer device, or to a  digital computer  for analysis. The popular Digital panel Meter (DPM) is a well known example of this. However, there are two major drawbacks in using it as a DAS. It is slow and the BCD has to be changed into  binary coding , if the output is to be processed by digital equipment. While it is free running, the data from the A/D converter is transferred to the interface register at a rat...

Concepts and types of Digital voltmeter

Digital Voltmeter (DVM) Definition : Digital Voltmeter is a  voltage-sensitive device . It  measures AC  or  DC voltage  and  displays  the value directly in  numeric form  instead of pointer deflection.  DVM  is an acronym for  Digital Voltmeter . DVM was first invented in  1954  by  Andrew Kay . There exist many factors that affect the measurement accuracy of a digital voltmeter(DVM). These are basically temperature, input impedance, variation in power supply voltage etc. As we know that an  analogue instrument  provides results by  pointer deflection  on a continuous scale. On contrary, a  digital instrument  provides results as  discrete numerals . Thus providing accuracy and versatility together. The input range of DVM may vary from  ±1 V  to  1000 V . Precision DVM offers  input resistance  of  1 GΩ  or high for a voltage range of less th...